Bridge ki default limit X-ray / DICOM file ke liye 60 MB aur 3D scan (STL / PLY / OBJ) ke liye 300 MB hai. Bridge Doctor ke Fix me "raise the size limit" likha aata hai, par X-ray Bridge Settings me limit badalne ka koi button nahi hai — zarurat ho to ORALZY support se karwana hoga. CBCT ek 3D scan me ~1000 projection frames aur dose report bhejta hai — ye X-ray nahi hain, isliye patient par nahi lagte. 3D scan ORALZY me chahiye to imaging software se volume / screenshot export karke daalo. Bridge Doctor naya feature hai — na dikhe to abhi aapki clinic ke liye chalu nahi hua (pehle beta clinics me aata hai).
Bridge Doctor me "skipped as too large" ya CBCT raw frames ka matlab?
"Files on <PC> are being skipped as too large" = kuch image bridge ki size limit se badi hain (OPG / CBCT DICOM me aam) — wo upload nahi hui. Ilaaj: X-ray software se sirf 2-D OPG export folder watch karo. "<PC> sends 3D (CBCT) raw frames — ORALZY ignores them" = sirf jaankari, kuch nahi karna.
Where in ORALZY: Setup → Integrations → Imaging Bridge (ORALINK) · Desktop: ORALZY menu → X-ray Bridge Settings…
Quick steps
- 🩺 Bridge Doctor me PC ka naam aur finding dekho
- Badi file wali finding ho to X-ray software me 2-D OPG export folder banao
- Wo folder X-ray PC par "X-ray Bridge Settings…" me jodo, bade DICOM wala folder ✕ se hatao → "Save"
- CBCT frames wali finding par kuch nahi karna