Autoclave ke validation tests kaunse chahiye — Bowie-Dick aur biological indicator kitni baar?

Daily: chemical indicator tape + Class IV strip har load + Bowie-Dick + leak rate. Weekly: biological indicator (spore vial). Har load Sterilisation Log me record.

Where in ORALZY: NABH → NABH → NABH Hub → Compliance Logs · Compliance Logs → "Sterilization Log" tab · NABH → NABH → SOP Library

Assessor validation records dekhta hai: har load me chemical indicator / Class IV strip, B-class autoclave par roz Bowie-Dick (aur leak test), aur hafte me biological indicator (spore test) ka result. ORALZY me har cycle NABH → NABH → NABH Hub → Compliance Logs → "Sterilization Log" tab → "Add Sterilization Log" me likho: Equipment Name, Cycle #, Cycle Result, Cycle Type, Temperature °C, Duration (min), Chemical Indicator, Biological Indicator (pass / fail / pending / na), Operator, Load Description. Ye "Autoclave Cycle Log" aur "Class 6 / TST Strip Test Register" me row bana deta hai; Bowie-Dick aur spore test ki entries NABH Registers ke apne register me hoti hain. Koi indicator fail ho to Cycle Result "fail" + "Action on failure" likho, poora load dobara process karo aur service bulao. Recall SOP: har pouch par batch no + date, fail hone par pichhle safal test ke baad ke loads wapas lo.

sterilizationbowie-dickbiological-indicatorspore-vialvalidationrecallsterilisation-logaudittaiyariautoclavecapaipc-6